CVE-2018-10601: ics-cert intellivue patient monitors, avalon fetal/maternal monitors Stack-based Buffer Overflow
Vulnerability summary
CVE-2018-10601 is a high (CVSS 8.2) Stack-based Buffer Overflow vulnerability in ics-cert intellivue patient monitors, avalon fetal/maternal monitors. EPSS estimates a 0.37% chance of exploitation in the next 30 days, higher than 29.6% of scored vulnerabilities.
Timeline
- 05-06-2018PublishedVulnerability first disclosed.
- 17-09-2024Last modifiedVulnerability information updated.
Description
IntelliVue Patient Monitors MP Series (including MP2/X2/MP30/MP50/MP70/NP90/MX700/800) Rev B-M, IntelliVue Patient Monitors MX (MX400-550) Rev J-M and (X3/MX100 for Rev M only), and Avalon Fetal/Maternal Monitors FM20/FM30/FM40/FM50 with software Revisions F.0, G.0 and J.3 have a vulnerability that exposes an "echo" service, in which an attacker-sent buffer to an attacker-chosen device address within the same subnet is copied to the stack with no boundary checks, hence resulting in stack overflow.
CVSS metrics
2 scores · 1 source · 2 versionsEPSS trend
Probability
0.37%
Chance of exploitation in the next 30 days
Percentile
30%
Versus all other CVEs
30-day change
0.00 pp
From 0.37% on 20-09-2026
Known exploited
Compare KEV catalogsNot in any of the 3 tracked catalogs.
Public exploits
0 foundNo public exploits indexed.
Weaknesses
CWE · root cause- CWE-121Stack-based Buffer OverflowA stack-based buffer overflow condition is a condition where the buffer being overwritten is allocated on the stack (i.e., is a local variable or, rarely, a parameter to a function).
- CWE-787Out-of-bounds WriteThe product writes data past the end, or before the beginning, of the intended buffer.
Affected systems
Philips18 products
- avalon_fetal\/maternal_monitors_fm20_firmwarena
- avalon_fetal\/maternal_monitors_fm30_firmwarena
- avalon_fetal\/maternal_monitors_fm40_firmwarena
- avalon_fetal\/maternal_monitors_fm50_firmwarena
- intellivue_mp2_firmwarena
Ics-cert1 product
- intellivue patient monitors, avalon fetal/maternal monitorsThe following IntelliVue Patient Monitors versions are affected: IntelliVue Patient Monitors MP Series (includingMP2/X2/MP30/MP50/MP70/NP90/MX700/800) Rev B-M, and IntelliVue Patient Monitors MX (MX400-550) Rev J-M and (X3/MX100 for Rev M only). The following Avalon Fetal/Maternal Monitors versions are affected: Avalon Fetal/Maternal Monitors FM20/FM30/FM40/FM50 with software Revisions F.0, G.0 and J.3
2 vendors · 19 products
References
1 links- ics-cert.us-cert.gov/advisories/ICSMA-18-156-01X Refsource MISCThird Party AdvisoryUS Government Resource