10

CVE-2018-4846: siemens ag rapidlab 1200 systems rapidpoint 400 systems rapidpoint 500 systems rapidlab 1200 series rapidpoint 500 systems rapidpoint 500 systems rapidpoint 500 systems rapidpoint 400 systems Use of Hard-coded Credentials

CVSS: 10.0v2.0HighModified
Published 26-06-2018Last modified 17-09-2024

Vulnerability summary

CVE-2018-4846 is a high (CVSS 10.0) Use of Hard-coded Credentials vulnerability in siemens ag rapidlab 1200 systems rapidpoint 400 systems rapidpoint 500 systems rapidlab 1200 series rapidpoint 500 systems rapidpoint 500 systems rapidpoint 500 systems rapidpoint 400 systems. EPSS estimates a 1.76% chance of exploitation in the next 30 days, higher than 77.6% of scored vulnerabilities.

Overall risk
70/100
0100
High · default modelHow is this calculated?
CVSS score
10.0/10
010
High · v2.0 (NVD)
EPSS score
1.76%
0%100%
Low · 0.00 pp in 30 days
KEVNot listedIn no KEV catalog
RansomwareNone recordedNo known use recorded
Public exploits0No public exploits recorded

Timeline

  1. 26-06-2018
    PublishedVulnerability first disclosed.
  2. 17-09-2024
    Last modifiedVulnerability information updated.

Description

A vulnerability has been identified in RAPIDLab 1200 systems / RAPIDPoint 400 systems / RAPIDPoint 500 systems (All versions_without_ use of Siemens Healthineers Informatics products), RAPIDLab 1200 Series (All versions < V3.3 _with_ Siemens Healthineers Informatics products), RAPIDPoint 500 systems (All versions >= V3.0 _with_ Siemens Healthineers Informatics products), RAPIDPoint 500 systems (V2.4.X_with_ Siemens Healthineers Informatics products), RAPIDPoint 500 systems (All versions =< V2.3 _with_ Siemens Healthineers Informatics products), RAPIDPoint 400 systems (All versions _with_ Siemens Healthineers Informatics products). A factory account with hardcoded password might allow attackers access to the device over port 5900/tcp. Successful exploitation requires no user interaction or privileges and impacts the confidentiality, integrity, and availability of the affected device. At the time of advisory publication, no public exploitation of this security vulnerability is known. Siemens Healthineers confirms the security vulnerability and provides mitigations to resolve the security issue.

CVSS metrics

2 scores · 1 source · 2 versions

EPSS trend

Probability

1.76%

Chance of exploitation in the next 30 days

Percentile

78%

Versus all other CVEs

30-day change

0.00 pp

From 1.76% on 20-09-2026

20-09-202620-10-2026
EPSS score
1.76%

Known exploited

Compare KEV catalogs

Not in any of the 3 tracked catalogs.

Public exploits

0 found

No public exploits indexed.

Weaknesses

CWE · root cause

Affected systems

Siemens3 products
Siemens ag1 product

2 vendors · 4 products

References

1 links