CVE-2019-10950: fujifilm cr-ir_357_fcr_capsula_x_firmware Improper Access Control
Vulnerability summary
CVE-2019-10950 is a high (CVSS 10.0) Improper Access Control vulnerability in fujifilm cr-ir_357_fcr_capsula_x_firmware. EPSS estimates a 3.63% chance of exploitation in the next 30 days, higher than 89.4% of scored vulnerabilities.
Timeline
- 30-04-2019PublishedVulnerability first disclosed.
- 04-08-2024Last modifiedVulnerability information updated.
Description
Fujifilm FCR Capsula X/ Carbon X/ FCR XC-2, model versions CR-IR 357 FCR Carbon X, CR-IR 357 FCR XC-2, FCR-IR 357 FCR Capsula X provide insecure telnet services that lack authentication requirements. An attacker who successfully exploits this vulnerability may be able to access the underlying operating system.
CVSS metrics
2 scores · 1 source · 2 versionsEPSS trend
Probability
3.63%
Chance of exploitation in the next 30 days
Percentile
89%
Versus all other CVEs
30-day change
0.00 pp
From 3.63% on 20-09-2026
Known exploited
Compare KEV catalogsNot in any of the 3 tracked catalogs.
Public exploits
0 foundNo public exploits indexed.
Weaknesses
CWE · root cause- CWE-284Improper Access ControlThe product does not restrict or incorrectly restricts access to a resource from an unauthorized actor.
- CWE-306Missing Authentication for Critical FunctionThe product does not perform any authentication for functionality that requires a provable user identity or consumes a significant amount of resources.
Affected systems
Fujifilm4 products
- cr-ir_357_fcr_capsula_x_firmwarena
- cr-ir_357_fcr_carbon_x_firmwarena
- cr-ir_357_fcr_xc-2_firmwarena
- fujifilm fcr capsula x/ carbon xCR-IR 357 FCR Carbon X | CR-IR 357 FCR XC-2 | FCR-IR 357 FCR Capsula X
1 vendor · 4 products
References
2 links- ics-cert.us-cert.gov/advisories/ICSMA-19-113-01X Refsource MISCThird Party AdvisoryUS Government Resource