CVE-2020-12023: philips intellibridge_enterprise
Vulnerability summary
CVE-2020-12023 is a medium (CVSS 4.5) vulnerability in philips intellibridge_enterprise. EPSS estimates a 0.52% chance of exploitation in the next 30 days, higher than 43.2% of scored vulnerabilities.
Timeline
- 11-06-2020PublishedVulnerability first disclosed.
- 04-06-2025Last modifiedVulnerability information updated.
Description
Philips IntelliBridge Enterprise (IBE), Versions B.12 and prior, IntelliBridge Enterprise system integration with SureSigns (VS4), EarlyVue (VS30) and IntelliVue Guardian (IGS). Unencrypted user credentials received in the IntelliBridge Enterprise (IBE) are logged within the transaction logs, which are secured behind the login based administrative web portal. The unencrypted user credentials sent from the affected products listed above, for the purpose of handshake or authentication with the Enterprise Systems, are logged as the payload in IntelliBridge Enterprise (IBE) within the transaction logs. An attacker with administrative privileges could exploit this vulnerability to read plain text credentials from log files.
CVSS metrics
3 scores · 2 sources · 2 versions- CVE.orgCNA2.0LowCVSS:3.1/AV:A/AC:H/PR:H/UI:N/S:U/C:L/I:N/A:NDiffers
EPSS trend
Probability
0.52%
Chance of exploitation in the next 30 days
Percentile
43%
Versus all other CVEs
30-day change
0.00 pp
From 0.52% on 21-09-2026
Known exploited
Compare KEV catalogsNot in any of the 3 tracked catalogs.
Public exploits
0 foundNo public exploits indexed.
Weaknesses
CWE · root causeAffected systems
Philips2 products
- intellibridge_enterprise≤ b.12
- intellibridge enterprise (ibe)< Versions B.12
1 vendor · 2 products
References
3 links- www.cisa.gov/news-events/ics-medical-advisories/icsma-20-163-01X Refsource MISC
- www.us-cert.gov/ics/advisories/icsma-20-163-01X Refsource MISCX TransferredThird Party AdvisoryUS Government Resource