7.2

CVE-2020-6977: ge ge ultrasound products Protection Mechanism Failure

Modified
Published 20-02-2020Last modified 04-08-2024

Vulnerability summary

CVE-2020-6977 is a high (CVSS 7.2) Protection Mechanism Failure vulnerability in ge ge ultrasound products. EPSS estimates a 0.43% chance of exploitation in the next 30 days, higher than 36% of scored vulnerabilities.

Overall risk
29/100
0100
Medium · default modelHow is this calculated?
CVSS score
7.2/10
010
High · v2.0 (NVD)
EPSS score
0.43%
0%100%
Low · 0.00 pp in 30 days
KEVNot listedNot listed by any tracked catalog.
RansomwareNot flaggedNo ransomware use recorded.
Public exploits0No public exploits recorded

Timeline

  1. 20-02-2020
    PublishedVulnerability first disclosed.
  2. 04-08-2024
    Last modifiedVulnerability information updated.

Description

A restricted desktop environment escape vulnerability exists in the Kiosk Mode functionality of affected devices. Specially crafted inputs can allow the user to escape the restricted environment, resulting in access to the underlying operating system. Affected devices include the following GE Ultrasound Products: Vivid products - all versions; LOGIQ - all versions not including LOGIQ 100 Pro; Voluson - all versions; Versana Essential - all versions; Invenia ABUS Scan station - all versions; Venue - all versions not including Venue 40 R1-3 and Venue 50 R4-5

CVSS metrics

2 scores · 1 source · 2 versions

EPSS trend

Probability

0.43%

Chance of exploitation in the next 30 days

Percentile

36%

Versus all other CVEs

30-day change

No change 0.00 pp

From 0.43% on 22-09-2026

22-09-202622-10-2026

Known exploited

Compare KEV catalogs
CISA KEV

Not in the CISA KEV catalog.

ENISA CNW

Not in the ENISA CNW catalog.

CIRCL

Not in the CIRCL catalog.

Public exploits

0 found

No public exploits indexed.

Weaknesses

CWE · root cause

Affected systems

Ge1 product
  • ge ultrasound productsVivid products - all versions | LOGIQ - all versions not including LOGIQ 100 Pro | Voluson - all versions | Versana Essential - all versions | Invenia ABUS Scan station - all versions | Venue - all versions not including Venue 40 R1-3 and Venue 50 R4-5

1 vendor · 1 product

References

1 links